Search the whole station

UC series
UC series
UC series

Heating Circulator

UC +50℃~300℃
  • Heating Power5.5kW~130kW
  • Power range6kW~136kW
  • Cooling power1.8kW~130kW
  • Temperture accuracy±0.5℃

WeChat 86 15152266993

Support outlet temperature control of thermal conductive media

Suitable for temperature control in more usage scenarios
Used with reactors, reactors, microchannels, distillation systems, etc., and also used for heating and cooling of other equipment. Widely used in pharmaceutical, chemical, biological, semiconductor, new materials and other industries

Model UC-5530N UC-7530N UC-A1030N UC-A1530N UC-A2530N UC-A3830N UC-A6030N UC-A9530N UC-A13030N
Temp  range 50℃~300℃
Controller PID adaptive controller
Temp control Heat-conducting medium outlet temp. control
Display/Record 7-inch color touch screen,recording temperature curve,data export using excel format 
Temp feedback Heat-conducting medium temperature feedback Pt100
Temp accuracy ±0.5℃
Heating power KW 5.5kW 7.5kW 10kW 15kW 25kW 38kW 60kW 95kW 130kW
Cooling
Capacity
kW at
300℃ 5.5kW 7.5kW 10kW 15kW 25kW 38kW 60kW 95kW 130kW
200℃ 5..5kW 7.5kW 10kW 13kW 25kW 38kW 60kW 95kW 130kW
100℃ 4kW 5kW 7kW 12kW 20kW 30kW 48kW 76kW 100kW
65℃ 1.8kW 2.2kW 3kW 5kW 8.5kW 12kW 19kW 30kW 40kW
Criculation pump
flow
Max35L/min 1.2BAR Max50L/min 1.2BAR Max50L/min 1.2BAR Max110L/min 1.5BAR Max150L/min 1.5BAR Max200L/min 1.5BAR Max250L/min 2.5BAR Max400L/min 2.5BAR Max400L/min 2.5BAR
Heater Flanged pipe electric heater
Independent temp. limite Lneya
Cooler Copper Tube Aluminum Fin air-cooled Condenser Plate heat exchanger/Tube heat exchanger
Safety Protection Self-diagnosis function; freezer overload protection; high pressure switch; overload relay; thermal protection device and
other safety protection
Closed Circulation
System
The whole system is a full closed circulation, there is No oil mist at high temperature,pressure do not rise up
when system is running. The system will supplement oil automatically at low temperature.
Connection size ZG3/4 ZG1 ZG1 ZG1 DN32 PN10 DN40 PN10 DN40 PN10 DN50 PN10 DN50 PN10
Dimension cm 40*70*130 45*70*160 45*70*160 70*100*175 70*100*175 80*120*185 150*100*185 200*145*205 200*145*205
Weight kg 110 120 160 195 265 395 615 1220 1400
Power max
AC 380V 50HZ
7kW 9kW 12kW 17kW 28kW 41kW 65kW 101kW 136kW
Case material Cold rolled sheet (REL7035)
Optional 220V 60HZ three phase,440V~460V 80HZ three phase

Model LY-5020 LY-A020 UC-5020 UC-A1020 UC-A1520
Temp  range 50℃~170℃
Controller PID adaptive controller
Tempcontrol Heat-conducting medium outlet temp. control
Temp feedback Heat-conducting medium temperature feedback Pt100
Temp accuracy ±0.5℃
Heating power KW 5.5kW 10kW 5.5kW 10kW 15kW
Cooling
Capacity
kW at
170℃     5.5kW 10kW 15kW
100℃     4kW 8kW 12kW
65℃     1.8kW 3.4kW 5kW
Criculation pump
flow
35L/min 1.5BAR 50L/min 1.5BAR 35L/min 1.5BAR 50L/min  1.5BAR 50L/min  1.5BAR
Heater Flanged pipe electric heater
Independent temp. limite Rainbow temperature protector
Cooler Copper Tube Aluminum Fin air-cooled Condenser Plate heat exchanger
Operation panel Display set temperature and measured temperature,key operation input temperature
Safety Protection overload relay、variety of security features such as thermal protection devices
Closed Circulation
System
The whole system is a full closed circulation, there is No oil mist at high temperature,pressure do not rise up
when system is running. The system will supplement oil automatically at low temperature.
Connection size ZG3/4 ZG3/4 ZG3/4 ZG3/4 ZG1
Dimension cm 40*70*130 40*70*130 40*70*130 40*70*130 70*100*175
Weight kg 70kg 80kg 78kg 128kg 170kg
Power max
AC 380V 50HZ
6kW 11kW 6kW 11kW 16kW
Case material Cold rolled sheet (REL7035)
Optional Optional 7-inch color touch screen,recording temperature curve,data export using excel format
Optional SUS304 shell
Optional 220V 60HZ three phase,440V~460V 80HZ three phase

Send us a message about UC series Heating Circulator
*
*
正在提交中….
提交成功!
提交失败!请稍后再试
邮箱错误!
电话错误!

RT +10℃~300℃

recommend products

loading…

已经是到最后一篇内容了!