Failures of Heating Cooling Circulators
The heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Related recommendations
-
Circulating Chiller for 100 mL Laboratory Reactor
1206When choosing the right circulating chiller for your 100 mL laboratory reactor, there are several factors to consider to ensure that the equipment can provide precise temperature control and efficient cooling. ...
View details -
Necessity of temperature control in fractional distillation of distillation
1445The pharmaceutical and chemical industry involves many chemical reactions, and with the continuous development of the pharmaceutical and chemical industry, there is a great demand for such instruments. The traditional open cooling and heating con...
View details -
Expansion valve description of cooling water circulation machine for electric vehicle parts testing
1488The demand for electric vehicle parts testing cooling water circulation machine in the testing of new energy vehicle battery industry is rising, so the choice is very important. Among them, the expansion valve is a more important accessory, and we...
View details -
Several Common Troubleshooting Methods for Ultra-low Temperature Freezer in Laboratories
16171. After power supply is switched on, the air switch of the main power supply is automatically disconnected.Elimination method: Check whether the compressor of refrigeration system is damaged or not, and replace the parts if damaged. After replaci...
View details