Wuxi Guanya Refrigeration Technology co., ltd

Several Common Troubleshooting Methods for Ultra-low Temperature Freezer in Laboratories

Laboratory ultra low temperature freezer is also called cryogenic storage box, cryogenic refrigerator, etc. Common - 40, - 60, - 86 degrees Celsius ultra low temperature freezers are suitable for cryogenic test of electronic devices and special materials, and for preservation of plasma, biomaterials, vaccines, reagents, tuna, etc. In addition, there are extreme ultra low temperature freezers of - 120 degrees Celsius, - 135 degrees Celsius and - 160 degrees Celsius, - 192 degrees Celsius. Although the refrigeration system and control circuit of ultra low temperature freezer in laboratory are complex, the common faults are mainly circulating system and circuit part.

Following are several common faults of ultra-low temperature refrigerators and their troubleshooting methods:


1. After power supply is switched on, the air switch of the main power supply is automatically disconnected. Elimination method: Check whether the compressor of refrigeration system is damaged or not, and replace the parts if damaged. After replacing the fittings, turn on the machine to check whether it is normal. The fault is usually caused by unstable power supply, because the * stage starts first, so the damage is usually in stage. In this case, as long as the delay regulated power supply is installed, the fault will not occur again.

2. Stop working before the temperature in the box reaches the set temperature. Exclusion: Although the temperature in the box has not reached the set temperature, the panel display temperature has reached the set temperature. This is due to the damage of the thermistor which detects the temperature and the need to replace the parts. If there is no original accessories, a set of temperature control system can be re-installed to eliminate the failure.

3. Poor refrigeration effect may be refrigerant leakage

4. If the ultra-low temperature refrigerator does not start in winter, it may be that the thermostat regulation value is too small.

We are a ultra low temperature freezer manufacturer. Welcome to choose our products.
Related News
Related Products
  • Water Cooled Chiller SLJ

    Water Cooled Chiller SLJ

    January 11, 2018Temperature ranges from - 5 ~ - 150 degrees, can meet different temperature; Adopt environmental protection refrigerants; Famous brand semi-closed piston compressor, semi-closed bipolar piston compressor, semi-hermetic screw compressor, main brand are BOCK.view
  • High-Low Temperature Test Chamber

    High-Low Temperature Test Chamber

    March 30, 2018Performance characteristics:1. In the refrigeration system, the condensation adopts air cooling mode, which has a wide temperature control range, which can meet the user's various needs.2. With a unique b...view
  • Humidity Test Chamber

    Humidity Test Chamber

    April 8, 2018Performance characteristics: 1. With wide temperature control range, can meet the user's various needs.2. With a unique balanced temperature control method, with stable and balanced heating capacity, high precision and high stable temperature control canview
  • Single Fluid Heat Transfer TCU SR

    Single Fluid Heat Transfer TCU SR

    April 10, 2018Working temperature range from -25℃to 200℃, other temperature range need to set to the widest temperature range(-50℃to 250℃) Centralized control of 2~20 reactors, each reactor's and jacket temperature can be set independently. One system control multipleview
  • Reactor Material TCU ZLF

    Reactor Material TCU ZLF

    April 10, 2018A set of cooling heat exchanger, heat exchanger, through proportional adjusting valve control cold heat into the heat exchanger, then through the uniform medium input to the reactor jacketed to do heat exchange and temp. Control, system has inside expansiview
  • Small Gas Condensing System

    Small Gas Condensing System

    April 10, 2018ModelHQ-0540HQ-1040HQ-2540HQ-5040HQ-0575WHQ-1075WHQ-2575WHQ-5075WGas flow5m3/h10m3/h25m3/h50m3/h5m3/h10m3/h25m3/h50m3/hLowest Temp.-40℃-40℃-40℃-40℃-75℃-75℃-75℃-75℃Primary Temp.There is no-20℃...view
products
Contact
Tel: + 86 13912479193
E-mail: sales@cnzlj.com
Add: No.203, Hongyun Road, Hongshan Street, Optoelectronics Park, Xinwu District, Wuxi City, Jiangsu Province, China
code chat