High temperature fault treatment of temperature shock test equipment

The whole function of Wuxi Guanya LNEYA thermal shock test chamber is computer controlled. It is a self-developed software with a good operation interface, which makes the operation and monitoring of the user simpler and more intuitive. The maintenance function can keep the running program in your operation. In the current state, the value of this block can be temporarily changed. The time parameter can be set on the screen to make the cooling, heating, and blue transfer switch automatically, according to the set value. The cold box and the hot box are independently controlled, and the box doors are independent of each other, and the use range of the test box is expanded (one box and three uses). The thermal shock effect of the temperature impact test equipment can be fully guaranteed. The test box door is interlocked with the circulating fan and the blue drive to protect the operator's safety. Once the door is opened, the power of the circulating fan and the blue drive will be automatically cut off.

The cause of the high-voltage failure of the temperature shock test equipment may be that the cooling water flow is insufficient and the rated water flow rate is not reached. The main performance is that the pressure difference between the unit inlet and outlet water becomes smaller (compared with the pressure difference at the beginning of the system operation), and the temperature difference becomes larger. The reason for the lack of water flow is that the system is short of water or air. The solution is to install an exhaust valve at the high point of the pipeline to exhaust; the pipeline filter is blocked or too fine, and the water permeability is limited. A suitable filter should be used. Clean the filter regularly; the pump is small and not compatible with the system.

The temperature impact test equipment condenser is fouled or clogged. Condensed water is generally tap water, it is easy to scale at 30 ° C or above, and because the cooling tower is open, it is directly exposed to the air, dust foreign matter can easily enter the cooling water system, causing the condenser to be blocked and the heat exchange area is small. It is inefficient and affects water flow. The performance is that the pressure difference between the inlet and outlet water of the unit is large, and the temperature difference is large. The temperature above and below the condenser is high, and the copper tube of the condenser is hot. The unit should be backwashed regularly and chemically cleaned and descaled if necessary.

The temperature shock test equipment has too much refrigerant charge. This situation generally occurs after maintenance, which is characterized by high suction and discharge pressure, equilibrium pressure, and high compressor operating current. It should be vented according to the suction and exhaust pressure and the equilibrium pressure and the operating current under rated conditions until normal. The temperature impact test equipment is mixed with non-condensable gases such as air and nitrogen in the refrigerant. This situation usually occurs after maintenance, and the vacuum is not complete. Can only be drained, re-vacuum, refill the refrigerant.

False alarm caused by electrical failure of temperature shock test equipment. Due to the moisture, poor contact or damage of the high voltage protection relay, the unit electronic board is damp or damaged, and the communication failure causes a false alarm. This kind of false fault, often the HP fault indicator on the electronic board is not bright or slightly bright, the manual reset of the high voltage protection relay is invalid, the computer displays “HPRESET”, or it disappears automatically, the compressor running current is normal, and the suction and exhaust pressure is also normal. .

If the high-voltage fault of the temperature shock test equipment occurs, it needs to be solved in time to avoid the big fault caused by the small fault and cannot be used in time.
Related News
  • LNEYA Is Exhibiting At Achema 2018

    LNEYA Is Exhibiting At Achema 2018

    January 11, 2018​The LNEYA Group will participate in the ACHEMA -World Forum and Leading Show for the Process Industries,Frankfurt am Main, Germany.view
  • Approaching LNEYA 5 Main Products

    Approaching LNEYA 5 Main Products

    January 11, 2018Industrial Chiller:Industrial Chiller varies from dynamic temp. Special Chiller:Special Chiller consist of Multi-reactor chiller (TCU).Environmental Chamber:The test object is connected to a testing adapter. Industrial Freezer:Industrial freezer consist oview
  • The Analysis Of Ultra Low Freezer'S Price

    The Analysis Of Ultra Low Freezer'S Price

    July 4, 2018What are the characteristics and functions of ultra low freezer: 1. Ultra low freezer can prevent the precision of machine equipment from being influenced by the change of oil temperature in the proce...view
  • The Application of Cold Plate Freezer

    The Application of Cold Plate Freezer

    July 4, 2018The refrigeration system of cold plate freezer adopts secondary super-cooler technology with the rapid refrigeration and low limiting temperature. It uses imported brand name semi-hermetic piston comp...view
  • 2018 Automotive Testing Expo

    2018 Automotive Testing Expo

    September 11, 2018Automotive Testing Expo 9.25-9.27view
  • The Analysis Of How To Maintain The Low Temperature Circulator In Daily Life

    The Analysis Of How To Maintain The Low Temperature Circulator In Daily Life

    July 4, 2018Guanya is a leading role in heating cooling circulator. It provides the customers with proposal and technology of controlling the cold and heat, temperature control system with stable property, extend...view
Related Products
  • SLJ -110℃~-150℃

    SLJ -110℃~-150℃

    January 11, 2018Temperature ranges from - 5 ~ - 150 ℃, can meet different temperature; Adopt environmental protection refrigerants; Famous brand semi-closed piston compressor, semi-closed bipolar piston compressor, semi-hermetic screw compressor, main brand are BOCK.view
  • Single Fluid Heat Transfer System SR

    Single Fluid Heat Transfer System SR

    April 10, 2018Working temperature range from -25℃to 200℃, other temperature range need to set to the widest temperature range(-50℃to 250℃) Centralized control of 2~20 reactors, each reactor's and jacket temperature can be set independently. One system control multipleview
  • Reactor Temperature System ZLF

    Reactor Temperature System ZLF

    April 10, 2018A set of cooling heat exchanger, heat exchanger, through proportional adjusting valve control cold heat into the heat exchanger, then through the uniform medium input to the reactor jacketed to do heat exchange and temp. Control, system has inside expansiview
  • LT -45℃~30℃

    LT -45℃~30℃

    April 10, 2018Low Temperature Cooling Chiller LT model: LT-5018N,LT-5040N,LT-5062N,LT-5090N,LT-50A1N,LT-50A2N; Medium Temperature Range: -45℃~30℃. ASET multifunctional controller, control outlet temperature of water tankview
  • TES -45℃~250℃

    TES -45℃~250℃

    February 28, 2019-45℃~+200℃ 2.5kW~25kW ±0.3℃view
  • TES -85℃~200℃

    TES -85℃~200℃

    February 28, 2019Temperature control range:-85 ° C ~ 200 ° CPower range:2.5kW~25kWTemperature control accuracy:±0.3°CModelTES-8525WTES-8555WTES-85A10WTES-85A15WTES-85A25WTemp. Range-85℃~200℃-85℃~200℃-85℃~200℃-85℃~200℃-85℃~200℃Heating P...view
products
Contact
Tel: + 86 18751546043
E-mail: sales@cnzlj.com
Add: No.203, Hongyun Road, Hongshan Street, Optoelectronics Park, Xinwu District, Wuxi City, Jiangsu Province, China
code chat