How to solve the problem of refrigeration pump circulator pump failure?
Cooling heating devices are required in many chemical pharmaceutical and laboratory applications. Then how to solve the failure of the refrigeration heating device is also concerned about how to solve it?
1. Circulating pump failure
The circulating pump fault is generally the thermal overload of the circulating pump. There are four main reasons for this fault: The first is that the outer circulating pipeline of the heat conducting oil is too long. At this time, it is necessary to shorten the outer circulation pipeline, reduce the bending of the outer circulation pipeline, and then click Equipment electrical cabinet surface thermal relay Reset button; the second is the circulation pipeline blockage, cooling and heating device which is good to tell everyone, check the circulation pipeline cleaning, install the filter, and then click the equipment electrical cabinet surface thermal relay Reset button; The third is that the circulating pump voltage is too low. It is recommended to measure whether the circulating pump voltage meets the specified range, improve the voltage environment, click the Reset button of the equipment electrical cabinet surface thermal relay; the fourth is that the thermal relay regulates the current too small. At this time, it is recommended to measure the pump current. Compare the range of thermal relays, Phillips screwdriver to properly rotate the thermal relay current, click on the equipment electrical cabinet surface thermal relay Reset button.
2. Sensor failure
A great possibility of sensor failure is wiring error. Once the fault is encountered, check the wiring situation. At this time, you should pay attention to two situations, one is sensor short circuit or temperature <-160 °C, and the other is sensor open circuit or temperature > 349 ° C.
If the sensor is short-circuited or the temperature is <-160 °C, it may be the fault of the sensor itself. At this time, it is recommended to replace the sensor (there is a spare sensor in the equipment slot). If the sensor is wired incorrectly, It is recommended to check the wiring (corrected circuit diagram) correction.

If the sensor is open circuit or the temperature is >349°C, it should be the fault of the sensor itself. It is necessary to replace the sensor in time (there is a spare sensor in the equipment slot). At this time, it is necessary to pay attention to whether the temperature is too high and the temperature is too high. Check whether the temperature is exceeded. Range, if the equipment is not powered, pay attention to check the power supply of the equipment.
These two faults of the cooling heating device are still relatively simple. I believe that everyone who has seen the sharing can solve it very quickly.
Raccomandazioni correlate
-
Application of refrigeration and heating temperature control system in chemical synthesis
866chemical synthesis cooling heating system temperature control unit
Visualizza i dettagli -
About vocs condensation adsorption machine resonance fault description
1101When choosing vocs condensation adsorption machine, LNEYA recommends selecting reliable vocs condensation adsorption machine equipment to avoid some simple faults, but if a similar resonance fault occurs, it needs to be solved in time. For VOCS ga...
Visualizza i dettagli -
Small environmental test chamber compressor should pay attention after maintenance
1078In the centrifugal small-scale environmental test chamber, the suction recovery device is generally an independent system, and if necessary, the pipeline connected to the condenser and the evaporator can be closed for separate maintenance. After r...
Visualizza i dettagli -
Cooling and shrinking assembly freezer for metal parts and common assembly methods in the industry
1522In the cold shrink production of copper sleeves, bearings, etc., quick-freezing cryogenic boxes are widely used in mechanical assembly. Copper sleeve cooling shrink assembly cabinet variable temperature assembly method is to heat the hole or cool ...
Visualizza i dettagli