Clean of Ultra-low Temperature Freezer
First wipe the machine shell from top to bottom with soapy water or disinfectant commonly used in the laboratory, then wipe it thoroughly with clean water, and dry it with a soft dry cloth.
Cautions: Avoid using too much water to clean the control board area, in order to prevent leakage damage of circuit components.
Cleanenvironment of ultra-lowtemperature freezer: Once a month
Clean up all ground debris and dust with sanitary ware, reduce dust adsorption in compressor and refrigeration system, affect heat dissipation
Prevent dust from adsorbing on circuit board, causing circuit components to be damaged by interference
Sealing test of door seals:
Once every six months, poor sealing will increase the working time of compressor and cause compressor failure.
Open the door, then put in a sheet of A4 paper, close the door, pull A4 paper out, if it is difficult to pull, it shows good sealing. It needs to be tested all around the door seal.
Clean air filter: Clean at least 4 times a year
Grasp the button at the bottom left corner and open the door directly below the cryogenic freezer.
Remove the black air filter and wash it with water or neutral detergent.
After drying, place the filter in place.
If it’s already dirty, the dust and debris on it can’t be removed, and a new filter needs to be replaced.
Clean condenser: at least twice a year
Refrigerator radiates heat to the outside environment through condenser.
Ash accumulation in condenser and filter will cause heat dissipation obstruction and decrease refrigeration efficiency of refrigerator.
1. Grasp the button at the bottom left corner and open the door directly below the refrigerator.
2. Use a vacuum cleaner to clean the dust on the condenser, or use compressed air to blow off the dust adsorbed on the condenser, or brush off the dust with a fine brush.
Notes: According to the environmental conditions of different laboratories, it is necessary to clean the condenser several times a year. And in the cleaning process, do not damage the radiator on the condenser.
Intracavity defrosting: once a year or when the thickness of defrosting exceeds 10 mm
High humidity, frequent door opening can easily lead to frost inside the door. Excessive frost, ice will affect the performance of refrigerators, can be deicing shovel and vacuum cleaner deicing.
1. Transfer samples from all chambers to another cryogenic refrigerator
2. Turn off the battery switch, the main power switch, and pull out the power cord.
3. Open all the inner doors and let the ice melt and wipe it off with a soft rag.
4. When the frost is removed, the inner cavity needs to be thoroughly cleaned with hot water or detergent (soda water) without chloride and wiped with a dry cloth.
5. Re-insert the power cord and turn on the main power switch and battery switch.
6. Samples can be placed after the refrigerator runs empty overnight.
We are a ultra low temperature freezer manufacturer. Welcome to choose our products.
Raccomandazioni correlate
-
What should do if the equipment of the thermal shock test chamber does not reach the set temperature
1225The choice of the manufacturer of the thermal shock test chamber is related to the quality of the thermal shock test chamber. Once the quality is set, the temperature setting of the equipment will not be closed. What should I do if I encounter the...
Visualizza i dettagli -
What are the functions of the reactor temperature control system?
966The reactor temperature control system is equipped with various reactors for high and low temperature reflection of liquid and solid. The temperature control system of LNEYA reactor adopts fully closed circulation pipeline, which is efficient in o...
Visualizza i dettagli -
Explosion-proof cryogenic refrigeration unit instructions
1029Explosion-proof cryogenic refrigeration units are used in many industries. When users use them, they need to understand their corresponding usage instructions to avoid improper operation and more efficient operation. The energy consumption of the ...
Visualizza i dettagli -
Il ruolo del sistema di controllo della temperatura TCU nel reattore chimico e farmaceutico
1268Il controllo della temperatura del bollitore di reazione nella ricerca e sviluppo di farmaci e negli esperimenti chimici, così come il controllo della temperatura nella produzione sperimentale su piccola scala e nei processi di produzione industriale, richiedono una tempera...
Visualizza i dettagli