Common Faults of Salt Spray Environmental Chambers

The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.

Salt spray environmental chamber for sale

The salt spray environmental chamber cannot be heated

1. Heat the solid state relay, if there is no output, replace the solid state relay.

2. If the external dial has no output, replace the external dial.

3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.

4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.

5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.


The heating temperature is abnormal

1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.

2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.

3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.

4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.


The external dial doesn't display

1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.

2. Check whether or not the power switch is damaged. If damaged, replace the power switch.

3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.

4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.

5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.

If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.

Related News
  • Brand Tips of Ultra Low Temperature Freezer

    Brand Tips of Ultra Low Temperature Freezer

    August 9, 2018When it comes to ultra low temperature freezer, a majority of people will get worried because ultra low temperature means the technology becomes more challenging. Compared with other refrigerant plant...view
  • High temperature test equipment related troubleshooting

    High temperature test equipment related troubleshooting

    October 10, 2018High-temperature test equipment can operate high-temperature test equipment more efficiently while maintaining low failure rate during operation, thus reducing the cost of the enterprise. If there is a related fault in Wuxi Guanya high-temperature test eqview
  • Motor test chiller compressor causes wear

    Motor test chiller compressor causes wear

    October 15, 2018Motor test chiller in operation need to pay attention to the compressor is a relatively complex component, once the motor test chiller compressor crankshaft wear and tear, it must be resolved in time, then, what are the reasons?view
  • Collocation Requirements For Lab Water Chiller

    Collocation Requirements For Lab Water Chiller

    July 4, 2018The lab water chiller is a relatively advanced cryopreservation equipment, which is suitable for blood stations, hospitals, health and epidemic prevention systems, scientific research institutes, elec...view
  • Suitable Freezer Manufacturer

    Suitable Freezer Manufacturer

    August 9, 2018For enterprise users, choosing a suitable freezer manufacturer is very important and must be carefully treated. Wuxi Guanya Refrigeration technology co., Ltd, as a professional freezer manufacturer, i...view
  • New energy vehicle testing comprehensive test box filled with heat transfer oil

    New energy vehicle testing comprehensive test box filled with heat transfer oil

    September 30, 2018In the process of running the new energy vehicle test comprehensive test box, it is necessary to pay attention to the use of the heat transfer oil filling of Wuxi Guanya before the operation. Then, how to increase the heat transfer oil of the new energy vview
x
Want some help from our team? LNEYA is dynamic temperature control equipment leader in China. We’ll contact you very soon.
Brilliant Features of LNEYA Dynamic Temperature Control System:
Contact us and we will offer the most suitable solution for you.
  • TEL:+ 86 510 82303782
  • FAX:+ 86 510 88587187
  • EMAIL:sales@cnzlj.com
  • ADDRESS:No.10, Zhenfa No.1 Road, Shuofang Street, New District, Wuxi, Jiangsu, China